ASTM E793 Standard Test Method for Enthalpies of Fusion and Crystallization by DSC

    What is ASTM E793? 

    ASTM E793 is a standard test method for determining the enthalpy of fusion (melting) and enthalpy of crystallization of materials using differential scanning calorimetry (DSC). This test method subjects the specimen to a heating or cooling cycle through its first-order transition while a DSC measures heat flow to or from the sample, and uses the area under the transition peak to determine the resulting enthalpy. ASTM E793 (Standard Test Method for Enthalpies of Fusion and Crystallization by Differential Scanning Calorimetry) is applicable to solid samples in granular form, or to any fabricated shape from which a specimen is cut, as well as to liquid samples that crystallize within the temperature range of operation. The standard applies to the normal operating temperature range of −120°C to 600°C, and technicians extend this range depending on the instrumentation they use.. Engineers rely on the resulting enthalpy values for quality control, specification acceptance, and research applications. 

    Read more

    Get Certified ASTM E793 Testing for Reliable Thermal Transition Performance 

    Certification confirms that a material’s enthalpy of fusion and crystallization meets ASTM E793 requirements, including uniform transition behavior between production batches. Manufacturers use ASTM E793-certified testing to determine a material’s thermal transition properties before accepting it for quality control or specification acceptance. Testing and certification ensure that the product meets the requirements of ASTM E793 and minimize the possibility of variable thermal performance in downstream applications. 

    What is the Scope of ASTM E793? 

    ASTM E793 is a test method that specifies how to use differential scanning calorimetry to measure the enthalpy of fusion and crystallization. It measures enthalpy changes associated with first-order transitions during heating and cooling of a sample through the melting or crystallization point. This standard offers a fast and uniform testing procedure to quantify these thermal transitions under controlled heating and cooling conditions with a DSC instrument.

    Read more

    The main objective of ASTM E793 is to determine the enthalpy of fusion and crystallization of thermally stable materials that exhibit well-defined transition behavior. This test is employed for quality control, specification acceptance, and in materials research in industry. 

    • Material type: Solid samples in granular form, or in any fabricated shape, and liquid samples that crystallize within the operating range of the instrument. 
    • Purpose: To determine the enthalpy of fusion and crystallization. 
    • Properties evaluated: Enthalpy (heat) of fusion and enthalpy of crystallization.
    • Testing Method: Heating or cooling of a specimen in a DSC instrument until it reaches the transition temperature to measure heat flow and to calculate the enthalpy value from the area of the transition peak. 
    • Result: The units of measurement for enthalpy are J/g. 
    • Test Limitations: In general, the test standard applies to thermally stable materials with well-defined endothermic or exothermic characteristics; results may vary with specimen shape and size.  
    • Use conditions: Enthalpy values measured are influenced by specimen mass, specimen shape, and heating or cooling rate. 
    • Applications: Polymer and material quality control, specification acceptance testing, thermal transition research, accredited materials testing. 

    What are the Uses of ASTM E793 Testing?

    ASTM E793 testing helps laboratories to establish the enthalpic variations associated with a material’s melting or crystallization transition. Engineers use these results to ensure quality control and to understand the fit of a material’s transition behavior to a set specification. This standard-

    • Compares the enthalpy of fusion of solid and liquid substances.
    • Manually alters the heating and cooling rates of a DSC instrument.
    • Calculates the value of the enthalpy of crystallization as a material cools through its transition point.
    • Helps ensure quality control throughout production runs.
    • Facilitates specification acceptance testing of thermally stable materials.
    • Helps with materials research and characterization of thermal behavior.
    • Supports selection of consistent, predictable transition behavior in materials 

    What Materials Can Be Tested Under ASTM E793? 

    ASTM E793 is applicable to solid samples in granular form or to any fabricated shape from which a suitable specimen can be cut by the technician. It also applies to liquid samples that crystallize within the instrument’s operating range. This test method is suitable for most thermally stable materials that have clearly defined endothermic or exothermic transitions; technicians select a specimen form and mass appropriate to the instrument and the transition under evaluation.

    Why is ASTM E793 Important? 

    ASTM E793 is significant as it is a quick and standardized method for measuring enthalpic variations associated with a material’s first-order thermal transitions. Manufacturers and researchers use these measured enthalpy values for quality control, specification acceptance, and materials research. 

    ASTM E793 Equipment and Sample Preparation Guide

    ASTM E793 measures heat flow from a differential scanning calorimeter as a sample goes through melting or crystallization. Technicians select a value of the specimen form and mass that is suitable for the material and the range of the instrument. The standard also allows for the use of computer- or electronic-based instruments, techniques, or data treatment equivalent to those described in the method. 

    Read more

    Sample and Specimen DetailsTechnicians use solid samples in granular form, fabricated shapes from which an appropriate specimen can be cut, or liquid samples that crystallize within the instrument temperature range.
    Specimen PreparationTechnicians prepare the specimen for DSC analysis, ensuring the specimen mass and form are appropriate for the material being tested. The standard notes that results may be affected by the form and mass of the specimen.
    Specimen DimensionsSpecimen size depends on the DSC instrument requirements; the standard specifies appropriate specimen mass for the analysis. 
    InstrumentationTechnicians use a differential scanning calorimeter (DSC) capable of operating from −120 °C to 600 °C, with a purge gas system (dry nitrogen or other inert gas at 10 to 50 mL/min), and a data acquisition system.

    Testing Procedures and Requirements for ASTM E793  

    The specimen is placed in the DSC instrument and heated or cooled over the temperature range that encompasses the melting or crystallization transition. The instrument is capable of continuously measuring heat flow and computing enthalpy from the area of the resulting transition peak. 

    Read more

    Calibration Technicians calibrate the heat flow axis of the instrument using the procedure in Practice E968, using the same heating rate, purge gas, and flow rate to be used for specimens.  
    Specimen Preparation and Loading Technicians prepare and load the specimen into the DSC instrument. They may heat the specimen rapidly to 50 °C below the expected melting temperature and allow it to equilibrate.
    Heating Scan Technicians heat the specimen at 10 °C/min through the melting range until the baseline is reestablished above the melting endotherm. Other heating rates may be used but must be noted in the report.
    Cooling Scan Technicians cool the specimen at 10 °C/min through the crystallization exotherm. Other cooling rates may be used but must be noted in the report.
    Result Reporting Laboratories report the enthalpy of fusion and crystallization, specimen mass, dimensions, geometry, material of the specimen capsule, and heating/cooling rates.

    ASTM E793 Testing Process and Data Collection

    Testing begins when the technician cuts and weighs the sample and places it in the DSC sample pan along with a reference pan. The technician then programs the heating or cooling rate and the temperature range for the melting or crystallization transition of the specimen. The instrument continuously measures heat flow as the specimen passes through its transition temperature. The software then calculates the area under the resulting transition peak. Lastly, the analyst uses this peak area to determine the enthalpy of fusion or crystallization, and reports the result in units of joules/gram. 

    The image shows a differential scanning calorimeter (DSC) being used to determine the enthalpies of fusion and crystallization of a material specimen according to ASTM E793.
    ASTM E793 Differential Scanning Calorimetry for Enthalpies of Fusion and Crystallization

    Common Challenges and Troubleshooting

    The results vary depending on specimen form and mass as stated by the standard, and inconsistent mass or form of the specimens often gives variable enthalpy readings. Poor thermal contact between the specimen and the sample pan also alters the heat flow measurement. Baseline drift during a long heating or cooling program complicates accurate peak integration. Laboratories solve these problems by using a weighing balance with a high degree of precision, good thermal contact of the specimen in the sample pan, and by checking baseline stability prior to each test. 

    ASTM E793 Analysis Results and Interpretation

    The laboratory gives the enthalpy of fusion or crystallization of the specimen of the material under test. ASTM E793 focuses on the enthalpic properties of a particular thermal transition rather than overall material performance.

    Read more

    • The laboratory records enthalpy of fusion and crystallization in joules per gram, calculated from the transition peak area.
    • The technicians determine the form and mass of the specimen tested, along with whether it is solid granular, fabricated shape, or liquid form.
    • The analyst makes note of the enthalpy measured at the given heating or cooling rate, correlated with the tested material and its transition behaviour under the mentioned heating/cooling rate. 
    • The engineer understands how to apply the results of the enthalpy measurement, taking into account the specimen mass, specimen shape, and heating or cooling rate.
    • Engineers use the measured enthalpy values to support quality control, specification acceptance, and materials research decisions. 

    Link to ASTM E793 

    FAQ

    What is the principle of ASTM E793?
    ASTM E793 measures enthalpic changes associated with first-order transitions of materials such as enthalpy of fusion (melting) and enthalpy of crystallization by differential scanning calorimetry (DSC).
    The normal operating temperature range is from −120 °C to 600 °C. The temperature range may be extended as needed, depending on the instrumentation.
    The method is used with solid samples that are in granular or fabricated form, or liquid samples where the liquid turns into a solid within the operating temperature range of the instrument.

    Updated on September 15, 2026

    Rohit Dhembare
    About Author
    Rohit Dhembare is an Operations Associate at MaTestLab Inc. and holds a postgraduate degree in Organic Chemistry.
    Testimonials
    Real results
    Engineers trust us with what matters most
    Start Your Testing
    Project Today
    Define your requirements and get access to
    specialized laboratories ready to deliver results
    Partners with us
    Clients
    Vendors
    Process for testing
    • STEP 01

      You share your testing requirements

    • STEP 02

      You share your sample(s)

    • STEP 03

      We deliver your test reports

    Get your testing done

    Let us known your testing requirements and we will be right back with a solution.

      Let us root for each other. Collaborate to grow, expand, and accelerate our businesses.

      Partner with us

        Discover more from Matestlab

        Subscribe now to keep reading and get access to the full archive.

        Continue reading