ASTM D6209 Standard Test Method for Determination of Gaseous and Particulate Polycyclic Aromatic Hydrocarbons in Ambient Air

    What is ASTM D6209?

    ASTM D6209 is a method for sampling, extracting, and analyzing the concentration of gaseous and particulate polycyclic aromatic hydrocarbons (PAHs) in ambient air using high-volume sampling, sample extraction, and gas chromatography/mass spectrometry (GC/MS). The method collects PAHs on a fine particle filter and a vapor trap filled with polyurethane foam (PUF) or XAD-2 resin. Ambient air is drawn at a rate of about 225 L/min through the sampling module for a user-selected time (1-24 h), but not more than 350 m³ of air is sampled. ASTM D6209 (Standard Test Method for Determination of Gaseous and Particulate Polycyclic Aromatic Hydrocarbons in Ambient Air (Collection on Sorbent-Backed Filters with Gas Chromatographic/Mass Spectrometric Analysis) outlines a standard method for measuring the combined concentration of gaseous and particulate-phase PAHs.

    Read more

    Get Certified ASTM D6209 Testing for Reliable Ambient Air PAH Evaluation

    ASTM D6209 testing is used to measure the concentration of PAHs in ambient air, where both particulate-associated and vapor-phase compounds are captured. The results may therefore be used to support ambient air monitoring, environmental assessments, regulatory evaluations, source investigations, and comparison of PAH concentration under controlled sampling conditions.

    What is the Scope of the ASTM D6209 Test Standard?

    ASTM D6209 outlines the procedures for sampling, cleanup, and analysis of PAH in ambient air. In the method, gaseous and particulate PAH are collected together in a high-volume sampling system, and the collected compounds are analyzed by GC/MS. It can be operated in the range of 100 to 250 L/min and can sample up to 350 m³ with a sampling period of up to 24 h, and has been tested for this. The method is applicable to a range of approximate concentrations of 0.05 to 1000 ng/m³ of sampled air. The scope of ASTM D6209 includes:

    Read more

    • Material: Ambient air containing gaseous and particulate-phase PAH.
    • Sampling method: High-volume collection through a fine-particle filter backed by a sorbent trap.
    • Sampling flow rate: 100 to 250 L/min, with approximately 225 L/min used in the described sampling procedure.
    • Sampling period: 1 to 24 h, depending on monitoring requirements and detection limits.
    • Maximum sampling volume: 350 m³ of air.
    • Collection media: A 104-mm particulate filter with a PUF or XAD-2 vapor trap.
    • Analysis: Combined extraction of the filter and sorbent followed by GC/MS analysis.
    • Measurement range: Approximately 0.05 to 1000 ng/m³ of sampled air.
    • Reporting basis: PAH concentrations are reported for the combined gas-phase and particulate-phase fractions.

    What are the Uses of ASTM D6209 Testing?

    ASTM D6209 is a controlled procedure for collecting and determining PAH in ambient air. Thus, the data can be used by technicians to aid in environmental monitoring, air-quality assessment, and investigations of PAH emissions from combustion sources. The ASTM D6209 standard-

    • Measures ambient air levels of gaseous and particulate PAHs.
    • Enables high-volume ambient air monitoring programs.
    • Assesses the levels of PAHs from combustion-related emission sources.
    • Delivers total (gas + particle) PAH measurement.
    • Supports environmental assessments and air quality assessments.
    • Supplies quantitative PAH concentration information for comparing sampling sites or times.
    • Supports regulatory and research programs that include ambient PAH.
    • Assists in the assessment of potential exposure to PAH compounds in the atmosphere.

    Which materials can be tested under ASTM D6209?

    ASTM D6209 can be used for ambient air samples where PAHs can be found in both the gaseous and particulate phases. The method has been tested for the PAHs included in its standard; other PAH may be determined if the user can show acceptable sampling and analytical efficiencies. The method utilizes a backup vapor trap, and the filter and sorbent extracts are combined to determine total PAH content because compounds with high vapor pressure may change between the particulate and gas phases during sampling.

    Why is ASTM D6209 Important?

    ASTM D6209 is important because PAH may be found in both the gaseous and particulate phases in ambient air,r and the distribution of PAH between these two phases will vary with temperature, humidity, particulate concentration, and residence time. The method offers a high-volume standardized method for sampling both phases and then calculating the concentration of both phases by GC/MS. The procedure also takes into account possible PAH loss from the particulate filter by volatilization and desorption during the sampling process.

    ASTM D6209 Equipment and Sample Preparation Guide

    ASTM D6209 uses a high-volume sampling module that has a 104-mm particulate filter and a sorbent cartridge. Technicians also require a controlled high-volume pump, flow-control and calibration equipment, pressure and temperature measurement devices, sample containers, extraction equipment, and a GC/MS system.

    Read more

    Sample and Specimen DetailsTechnicians collect representative ambient air through a 104-mm micro-quartz-fiber particulate filter backed by a PUF or XAD-2 sorbent trap.
    Specimen PreparationAnalysts clean and prepare sampling media, load the sorbent cartridge, assemble the filter and sorbent cartridge under clean conditions, and do necessary blank checks prior to field use.
    Specimen DimensionsThe particulate filter is 104 mm in diameter; the PUF sorbent is 76 mm long × 62 mm diameter, while the XAD-2 bed is approximately 5 cm deep and requires 55 to 60 g of resin.
    InstrumentationThe equipment consists of a high-volume pump of 250 L/min, a flow-control calibration device, a manometer, a thermometer and a barometer, a Soxhlet extractor, concentrators, a vacuum oven, and a GC/MS equipped with appropriate GC columns.

    Testing Procedures and Requirements for ASTM D6209

    As per ASTM D6209, technicians first calibrate the sampling system, then prepare the filter and sorbent cartridge, collect ambient air at the desired flow rate, preserve collected media, extract the filter and sorbent in the same cartridge, concentrate the extract, and analyze the sample using GC/MS. Care must be taken in the sampling flow, contamination, handling, and extraction conditions in this method. The procedure outlined in ASTM D6209 is as follows-

    Read more

    Sampling System CalibrationTechnicians calibrate the sampler flow-control system and verify the flow rate before and after sample collection using suitable calibrated flow-measurement equipment.
    Sampling Media PreparationAnalysts prepare and clean the particulate filter and PUF or XAD-2 sorbent, assemble the sampling cartridge, and verify acceptable blank levels before field use.
    Sampling Module AssemblyTechnicians install the sorbent-filled cartridge and particulate filter in the sampling module, tighten the connections, and check the system for proper sealing.
    Ambient Air CollectionOperators draw ambient air through the filter and vapor trap for 1-24 hours at about 225 L/min, but no more than 350 m³ of air is sampled in total.
    Sample Recovery and ExtractionTechnicians take the filter and sorbent cartridge out under clean conditions after sampling, transport them under controlled conditions, and extract them simultaneously in a Soxhlet system.
    Result ReportingAnalysts extract the sample and then inject an aliquot of the extract into the GC/MS to calculate the total concentration of the PAH of interest in the air sample.

    ASTM D6209 Testing Process and Data Collection

    ASTM D6209 uses a high-volume sampler that pulls ambient air through a fine-particle filter, and then through a PUF or XAD-2 vapor trap. The sampling site, sampling time, flow data, and volume of air sampled are recorded by the operators. The filter and sorbent are then removed by the analyst,t and the extract is concentrated and then subjected to GC/MS analysis. The data shown are the total gas- and particle-phase concentrations of the PAH measured.

    ASTM D6209 ambient air sampling module with particulate filter and sorbent trap connected to a high-volume sampling system.
    ASTM D6209 high-volume sampling system is used to collect gaseous and particulate PAH from ambient air.

    Common Challenges and Troubleshooting

    ASTM D6209 common challenges include contamination of sampling media, air leaks in the sampling module, inaccurate flow measurement, PAH loss through volatilization, inadequate sorbent preparation, and interference during GC/MS analysis. Therefore, it is important that the sampling components are cleaned, flow calibration is checked, there are proper seals, sample handling is controlled, and appropriate blanks are analyzed before the results are accepted.

    ASTM D6209 Analysis Results and Interpretation

    The results of ASTM D6209 are reported as concentrations of PAH in the ambient air sampled, typically in ng/m³. For this reason, the measured PAH concentration is combined with the volume of air sampled and the sampling conditions to assess air contamination in the environment. Combined gas phase and particulate phase results are obtained because the separation of the particulate phase and gas phase in the filter and vapor trap does not accurately reflect the atmospheric phase distribution.

    Read more

    • The laboratory documents the sampling site, the sampling time, the start of the sample, and the amount of air sampled.
    • The analyst determines the concentrations of the targeted PAH from the GC/MS analysis.
    • The laboratory indicates the concentration of PAH in ng/m³ of sampled air.
    • Engineers compare the concentration of PAHs at sampling locations or sampling times under known conditions.
    • Analysts use these results to help support ambient air monitoring, environmental assessment, regulatory evaluation, and research programs.

    Link to ASTM D6209

    FAQ

    Why is ASTM D6209 Important?
    The importance of ASTM D6209 is that it offers a standard method for measuring gaseous and particulate PAH levels in ambient air.
    ASTM D6209 can also be used to collect and determine PAH in ambient air for air-quality monitoring, environmental assessments, regulatory evaluations, and research.
    ASTM D6209 quantifies the levels of specific gaseous and particulate PAH in ambient air, and the results are typically expressed in ng/m³.

    Updated on September 23, 2026

    Malhar Khole
    About Author
    Malhar Khole is a Material Testing Associate at Matestlab Inc. and holds a postgraduate degree in Material Science and Technology from IIT (BHU), Varanasi.
    Know More
    Testimonials
    Real results
    Engineers trust us with what matters most
    Why Choose Matestlab?
    Matestlab is a unified testing services partner based in the USA. Matestlab offers raw material and product testing services through accredited, expert laboratories primarily based in the USA. Matestlab supports physical, thermal, chemical, electrical, mechanical, environmental, and characterization testing needs through its global network of 2200+ accredited testing laboratories. Matestlab's in-house materials science experts curate and design comprehensive testing programs matched to the sample material and specific testing need, with regulatory requirements and confidentiality in mind.

    If you are looking for a managed and trusted testing partner to achieve your research goals, feel free to schedule a meeting with us or send us a request at connect@matestlabs.com to learn more about our services. For a quick quote, call us at (302)244-7710. We are happy to serve you.
    Start Your Testing
    Project Today
    Define your requirements and get access to
    specialized laboratories ready to deliver results
    Partners with us
    Clients
    Vendors
    Process for testing
    • STEP 01

      You share your testing requirements

    • STEP 02

      You share your sample(s)

    • STEP 03

      We deliver your test reports