The result is an accuracy class for the extensometer system. ASTM bases the class on the maximum error of strain indicated. Therefore, readers should read the class together with the extensometer type. The points below explain how to read the results.
- The laboratory reports the assigned accuracy class, including Class A, B-1, B-2, C, D, or E, based on the applicable maximum permissible strain-measurement error. Analysts use this classification to assess the extensometer system’s accuracy.
- Analysts identify the extensometer type alongside its accuracy class. Types 1, 2, and 3 describe the applicable gauge-length categories, so reporting both designations provides a clearer description of the calibrated system.
- Technicians document gauge length in mm or inches, calibration measurements, indicated strain, reference displacement, and calculated measurement errors. They record calibration conditions and equipment identification to support traceability.
- Engineers evaluate whether the assigned accuracy class meets the intended test’s measurement requirements. More stringent accuracy requirements may apply when determining elastic modulus or other properties that depend on precise strain measurements.
- Analysts report strain as a dimensionless quantity and maintain consistent calibration procedures regardless of whether displacement measurements use SI or US customary units. Analysts consider the applicable class and extensometer configuration when interpreting measurement reliability.
FAQs
What does ASTM E83 cover?
ASTM E83 covers the calibration, verification, and classification of extensometer systems based on their strain measurement errors. It helps determine whether an extensometer meets the accuracy requirements for its intended use. The practice supports reliable strain measurements during material testing.
What classes does ASTM E83 define?
ASTM E83 defines Classes A, B-1, B-2, C, D, and E based on maximum permissible strain measurement errors. It also classifies extensometer systems into Types 1, 2, and 3 according to gauge length. These classifications help users identify the appropriate system for their measurement requirements.
Which instruments does ASTM E83 cover?
ASTM E83 covers instruments that indicate or record values proportional to changes in length during tensile or compressive strain measurements. It does not cover bonded resistance strain gauges attached directly to a specimen. Users should check the practice’s scope to confirm whether their instrument qualifies.
What factors affect the cost of extensometer verification?
The cost of extensometer verification depends on the number of instruments, their gauge lengths, and the required reporting details. Different verification requirements may also affect the overall cost. The testing laboratory can provide a quote based on the instruments and services required.
How long does extensometer verification take?
The turnaround time depends on the number of extensometers, the verification requirements, and laboratory availability. Instrument configuration and reporting needs may also affect the schedule. Confirm the expected completion time with the testing laboratory before submitting the instruments.
Direct Link to ASTM E83 PDF