ASTM E1695 Standard Test Method for Measurement of Computed Tomography (CT) System Performance

    What is ASTM E1695?

    ASTM E1695 is a standard test method for measuring the performance of X-ray and gamma-ray computed tomography (CT) systems, and it determines spatial resolution and contrast sensitivity from the CT volume of a uniform cylinder of material. The method derives spatial resolution, expressed as the modulation transfer function (MTF), from the sharpness at the edges of the reconstructed cylinder slices. It derives contrast sensitivity, expressed as the contrast discrimination function (CDF), from the contrast and the statistical noise at the center of the slices. This standard is important for CT inspection because it lets purchasers and providers of CT systems or services measure and specify performance, which in turn supports acceptance testing and quality assurance. 

    Read more

    Get Certified ASTM E1695 Testing for Reliable CT Performance Data

    Manufacturers, CT system providers, healthcare equipment developers, and industrial inspection teams use ASTM E1695 testing from a qualified, accredited laboratory to get measured CT performance data. The lab scans a uniform cylinder and reports spatial resolution as the modulation transfer function (MTF) and contrast sensitivity as the contrast discrimination function (CDF). Both results shift with the scan setup, so the report lists the acquisition parameters, reconstruction settings, and reconstructed volume details beside the numbers, along with any observations from the analysis.

    What is the Scope of ASTM E1695?

    ASTM E1695 is a standard test method for determining the spatial resolution and contrast sensitivity of X-ray and gamma-ray computed tomography (CT) volumes; results are expressed as the modulation transfer function and the contrast discrimination function, and the test uses a uniform cylinder of material.

    Read more

    The scope of the test method includes:

    • Test type: Measurement of CT system performance from the CT volume of a uniform cylinder
    • Radiation: X-ray and gamma-ray CT systems
    • Spatial resolution: Quantified as the modulation transfer function (MTF) from the sharpness at the edges of the reconstructed slices
    • Contrast sensitivity: Quantified as the contrast discrimination function (CDF) from the contrast and statistical noise at the center of the slices
    • Purpose: A way for the purchaser or provider of CT systems or services to measure and specify performance
    • Scan settings: Technicians record the acquisition settings, because the results depend on them

    What are the Uses of This Test Method?

    The method gives purchasers, providers, and users of CT systems a common way to measure image quality. The uses include:

    • Measures the spatial resolution of a CT system as the modulation transfer function.
    • Measures the contrast sensitivity of a CT system as the contrast discrimination function.
    • Lets a purchaser and a provider agree on measurable performance before they buy or sell a system or a service.
    • Supports acceptance testing and periodic checks of CT systems.
    • Supports comparison of systems and scan settings with the same test object.

    What systems can be tested under ASTM E1695?

    The method applies to X-ray and gamma-ray computed tomography (CT) systems that produce CT volumes. It covers the entire system, from the radiation source and detector to the reconstruction process, rather than evaluating individual components. The 2020 revision updated the terminology from images to volumes to accommodate modern CT systems that reconstruct complete three-dimensional volumes as well as slice-based images. ASTM E1695 supports industrial non-destructive testing(NDT) applications by providing a standardized approach to evaluating CT image quality and system performance without damaging the test object.

    Why is ASTM E1695 important?

    A CT image only helps an inspector if the system resolves small features and separates small differences in density. The method turns those two abilities into numbers, the MTF and the CDF, so buyers and sellers can specify them and check them. Because the test uses one uniform cylinder, results from different systems and settings are easy to compare. The test method falls under ASTM Committee E07 on Nondestructive Testing, Subcommittee E07.01 on Radiology (X and Gamma) Method.

    ASTM E1695 Equipment and Sample Preparation Guide

    The method uses a CT system, a uniform cylinder of material, and image analysis software. Technicians scan the cylinder under the desired settings and analyze the reconstructed slices.

    Read more

    Sample and Specimen DetailsThe test sample is a uniform cylinder of material that serves as the test object. The material is uniform so that the edge and the center of each slice give clean measurements.
    Specimen PreparationThe technician places the cylinder in the CT system and sets the acquisition parameters to the conditions the report will describe. The technician records those parameters.
    Specimen DimensionsThe cylinder is uniform in material and sized for the system. Its diameter spans enough voxels in the reconstructed slice, and its height covers the part of the detector used for the measured slices.
    InstrumentationThe laboratory uses the CT system with its source and detector, reconstruction software, and image analysis software that calculates the MTF from the edge of the cylinder and the CDF from the center.

    Testing Procedures and Requirements for ASTM E1695

    Technicians scan the uniform cylinder under the chosen settings, reconstruct the volume, and select slices through it. Analysts measure edge sharpness for spatial resolution and center contrast and noise for contrast sensitivity, then report both results with the scan settings. The ASTM E1695 testing procedure is as follows:

    Read more

    Test Object SetupTechnicians mount the uniform cylinder in the CT system so that it sits in the field of view and stays stable during the scan.
    Scan Parameter SetupTechnicians set the acquisition parameters to the conditions that they want to describe and record.
    Volume Acquisition & ReconstructionTechnicians scan the cylinder and reconstruct the CT volume.
    Slice SelectionAnalysts select the reconstructed slices through the cylinder that they use for the measurements.
    Spatial Resolution AnalysisAnalysts measure the sharpness at the edges of the cylinder in each slice and calculate the modulation transfer function (MTF).
    Contrast Sensitivity AnalysisAnalysts measure the contrast and the statistical noise at the center of the cylinder in each slice and calculate the contrast discrimination function (CDF).
    ReportingThe laboratory reports the MTF and CDF results together with the scan settings that influence them.

    ASTM E1695 Testing Process and Data Collection

    Technicians begin testing by mounting the uniform cylinder in the CT system and setting the acquisition parameters. They scan the cylinder, reconstruct the volume, and select slices through it. Analysts then measure the sharpness at the edges of the cylinder and calculate the modulation transfer function. They also measure the contrast and the statistical noise at the center and calculate the contrast discrimination function. The laboratory records the scan settings, the slices used, and both results.

    The image shows the ASTM E1695 testing procedure, including test object setup, scan parameter configuration, CT volume acquisition, slice selection, MTF and CDF analysis, and performance result reporting.
    ASTM E1695 Testing Procedures for Measuring Computed Tomography (CT) System Performance.

    Common Challenges and Troubleshooting

    The results follow the scan settings, so the first job is making sure the scan matches the settings written in the report. The cylinder also has to be uniform, because any variation in the material throws off both the edge and the center readings. Noise at the center lowers contrast sensitivity, which is why analysts only compare CDF values from the same cylinder scanned under similar settings. Reconstruction artifacts near the edge can blur the sharpness measurement, so technicians look through the slices before analyzing them. If the cylinder sits off center or shifts during the scan, the edge data becomes unreliable, and a firm mount prevents that.

    ASTM E1695 Analysis Results and Interpretation

    This report summarizes how the CT system performed in resolving fine details and distinguishing small differences in contrast under the selected scanning conditions.

    Read more

    • The laboratory reports spatial resolution through the modulation transfer function (MTF) and contrast sensitivity through the contrast discrimination function (CDF) for the reconstructed cylinder slices.
    • Technicians document the scan settings, acquisition parameters, reconstructed volume, and selected slices to support accurate review and reproducibility.
    • Analysts examine the edge sharpness of the cylinder and evaluate contrast and statistical noise at the center of each slice.
    • Engineers interpret the MTF and CDF results to assess CT system performance under the specific test conditions.
    • Purchasers and providers use these findings to compare CT systems, verify agreed performance requirements, and support acceptance testing and quality assurance.

    FAQs

    What does the method cover?

    The method covers the measurement of spatial resolution and contrast sensitivity of X-ray and gamma-ray computed tomography (CT) volumes.

    What test object does the method use?

    The method uses the CT volume of a uniform cylinder of material.

    How does the method measure spatial resolution?

    It derives the modulation transfer function from the sharpness at the edges of the reconstructed cylinder slices.

    How does the method measure contrast sensitivity?

    It derives the contrast discrimination function from the contrast and the statistical noise at the center of the cylinder slices.

    Link to ASTM E1695 Standard

    Updated on November 24, 2026

    Laraib Hashmi
    About Author
    Laraib Hashmi is an aspiring research enthusiast and science content professional with a strong interdisciplinary skill set, holding a postgraduate degree in Applied Microbiology from KIIT Bhubaneshwar, Odisha.
    Know More
    Testimonials
    Real results
    Engineers trust us with what matters most
    Why Choose Matestlab?
    Matestlab is a unified testing services partner based in the USA. Matestlab offers raw material and product testing services through accredited, expert laboratories primarily based in the USA. Matestlab supports physical, thermal, chemical, electrical, mechanical, environmental, and characterization testing needs through its global network of 2200+ accredited testing laboratories. Matestlab's in-house materials science experts curate and design comprehensive testing programs matched to the sample material and specific testing need, with regulatory requirements and confidentiality in mind.

    If you are looking for a managed and trusted testing partner to achieve your research goals, feel free to schedule a meeting with us or send us a request at connect@matestlabs.com to learn more about our services. For a quick quote, call us at (302)244-7710. We are happy to serve you.
    Start Your Testing
    Project Today
    Define your requirements and get access to
    specialized laboratories ready to deliver results
    Partners with us
    Clients
    Vendors
    Process for testing
    • STEP 01

      You share your testing requirements

    • STEP 02

      You share your sample(s)

    • STEP 03

      We deliver your test reports

    Get your testing done

    Let us know your testing requirements and we will be right back with a solution.

      Let us root for each other. Collaborate to grow, expand, and accelerate our businesses.

      Partner with us