ASTM E1064 Standard Test Method for Water in Organic Liquids by Coulometric Karl Fischer Titration

    What is ASTM E1064? 

    ASTM E1064 describes a standard test method that specifies a Karl Fischer reagent and an automated coulometric titration procedure to determine water in the range of 0 to 2.0% mass in most liquid organic chemicals. This test method involves adding an aliquot of the sample to the titration vessel of a coulometric Karl Fischer apparatus, which produces iodine at the anode that reacts with any water in the sample. Laboratories using ASTM E1064 (Standard Test Method for Water in Organic Liquids by Coulometric Karl Fischer Titration) rely on this coulometric technique because it suits determining low concentrations of water in organic liquids that would yield only small, less precise titers using the volumetric Karl Fischer titration procedure in Test Method E203. This test method is not applicable to liquefied products, such as liquefied petroleum gas (LPG), butane, propane, or liquid natural gas (LNG). Engineers compare results with the moisture purity required for organic liquid chemicals in manufacturing and quality control processes sensitive to moisture.

    Read more

    Get Certified ASTM E1064 Testing for Reliable Trace Water Content Verification 

    Certification confirms that a liquid organic chemical meets the requirements of the applicable ASTM E1064, including trace water (within specified purity limits). ASTM E1064 certified testing units test moisture content in a batch of solvent or organic liquid before accepting the batch for moisture-sensitive manufacturing or quality control. Testing and certification confirm compliance with ASTM E1064 requirements and minimize the risk of moisture contamination of the process. 

    What is the Scope of ASTM E1064 Standard? 

    ASTM E1064 provides a method for the determination of water content in the range of 0% to 2.0% mass in most liquid organic chemicals using the Karl Fischer reagent method. It uses an automated coulometric titration procedure for evaluating trace water content. This standard establishes a uniform test method for electrochemical titration of low-level water concentrations in organic liquids under controlled conditions using a coulometric Karl Fischer apparatus.

    Read more

    ASTM E1064 is used to determine the amount of water in an organic liquid by oxidizing the water in the sample with electrochemically generated iodine. This test is employed in industries to certify chemical purity, for quality control, and to monitor moisture-sensitive processes.

    The scope of ASTM E1064 includes 

    • Material type: Liquid organics (not liquefied gas products (LPG, butane, propane, LNG)). 
    • Purpose: Quantification for evaluation of trace water content. 
    • Properties measured: Water content (as % mass). 
    • Method: A sample aliquot is injected into a Karl Fischer titration vessel with electrochemically generated iodine to react with the water in the sample. 
    • Result: Water content of 0% to 2.0% mass is measured. 
    • Test Limitations: This test method is applicable to most liquid organic chemicals, but not applicable to liquefied gas products, for which test method E203 is applicable. 
    • Environmental conditions: Reagent purity, moisture contamination from the atmosphere, and injection technique may impact the measured water content. 
    • Applications: Chemical purity check, solvent quality control, moisture-sensitive process monitoring, low water concentration measurement. 

    What are the Uses of ASTM E1064 Testing?

    ASTM E1064 testing is a highly sensitive coulometric titration method that is used to determine trace water in organic liquids for laboratories. Chemical purity is then confirmed by quality control teams using the results, and moisture-sensitive manufacturing processes are then monitored. This standard-

    • Determines trace water levels in organic liquid chemicals.
    • Electrochemically produces iodine, which can be used to react with the water in a sample.
    • Evaluates low water concentrations that volumetric titration methods are not precise in measuring.
    • Helps ensure purity of the solvents used in moisture-sensitive manufacturing processes.
    • Facilitates batch-to-batch comparison of water content in organic liquids.
    • Helps with quality control in organic chemical manufacturing.
    • Helps select the Karl Fischer method appropriate to the range of water concentration in the samples. 

    What Materials Can Be Tested Under ASTM E1064? 

    ASTM E1064 is applicable for most liquid organic chemicals that contain 0% to 2.0% mass of water. Typical applications include solvents, industrial organic chemicals, and process liquids where moisture content directly affects product quality or downstream reactions. This test method is not applicable to liquefied gas products (LPG, butane, propane, LNG) that require a different approach to analysis. 

    Why is ASTM E1064 Important? 

    ASTM E1064 provides highly sensitive detection of trace water content that volumetric Karl Fischer titration alone cannot detect. Manufacturers use this test method because the presence of even a small quantity of water may affect moisture-sensitive chemical reactions or may serve as a sign of contamination during storage and handling. 

    ASTM E1064 Equipment and Sample Preparation Guide

    ASTM E1064 uses a coulometric Karl Fischer titration apparatus to produce electrochemically prepared iodine that reacts with water in the sample. A small sample of the organic liquid is directly pipetted into the titration flask. The standard specifies that this is a “special” coulometric method for low water concentrations as compared to the volumetric approach applied in Test Method E203 for higher water concentrations.

    Read more

    Sample and Specimen DetailsTechnicians test liquid organic chemical samples, excluding liquefied gas products, prepared for controlled coulometric Karl Fischer titration. 
    Specimen PreparationSamples are introduced using a syringe with a long needle immersed beneath the reagent surface. The syringe is weighed before and after injection to determine actual sample weight.
    Specimen DimensionsSample size depends on expected water content. For 1000 ppm water, 0.1–1 g sample is recommended; for 10 ppm, 5 g is recommended.
    InstrumentationThe test uses a coulometric Karl Fischer titrator with a generator electrode, indicator electrode, titration cell, and reagent solutions.

    Testing Procedures and Requirements for ASTM E1064  

    A sample aliquot is added to the titration vessel of the coulometric Karl Fischer apparatus, where the apparatus produces iodine electrochemically at the anode. The instrument determines the amount of water based on the amount of electrical charge that is needed to produce the iodine, which reacts with the water in the sample. Test Method E203 offers an alternative volumetric Karl Fischer method for higher water content, and ASTM D6304 refers to this test method for water in petroleum products, petroleum-based lubes and additives. 

    Read more

    Initial Setup Technicians turn on the coulometric titrator and allow the instrument to titrate ambient moisture in the reagent/cathode cell until a stable drift is achieved. 
    Sample Injection After flushing the syringe with sample fluid, technicians draw the sample slightly beyond the desired volume, invert the syringe to expel air bubbles, and inject the sample into the reagent through the septum. 
    Electrochemical Iodine Generation The apparatus generates iodine electrochemically at the anode, which reacts with the water present in the injected sample. 
    Charge Measurement The coulometer measures the electrical charge consumed during iodine generation, which corresponds directly to the water content present. 
    Result Reporting The instrument displays the water content value, which is recorded by date and time. Laboratories report results in SI units as mass percent or ppm.

    ASTM E1064 Testing Process and Data Collection

    Testing starts once the technician withdraws a small sample of the organic liquid with a syringe, allowing little exposure to atmospheric moisture while handling. This aliquot is then added to the titration vessel of the coulometric Karl Fischer apparatus. The apparatus then electrochemically produces iodine at the anode, which reacts with the water in the sample via the Karl Fischer reaction. During this iodine generation process, the electrical charge consumed is continuously measured in the coulometer. Lastly, the water content is computed as a percent mass of the total charge measured and reported to be between 0% and 2.0% mass. 

    Common Challenges and Troubleshooting

    Moisture contamination of the atmosphere often occurs and causes errors, as even a few seconds of exposure of the sample or the apparatus to atmospheric moisture can result in adding water that is not part of the organic liquid itself. If the Karl Fischer reagent is not fresh or stored properly, it can also lead to reagent degradation, which can impact measurement accuracy. If the technician is not careful and quick in drawing and injecting the aliquot, there may be some variability due to inconsistent technique. Laboratories solve these problems by reducing exposure of samples to open air, checking reagent freshness, and training technicians in standardized injection technique. 

    ASTM E1064 Analysis Results and Interpretation

    This report summarizes the trace water content found in the tested organic liquid sample. It displays the values measured, the sample data, and the conditions of the titration performed during the test. This interpretation is a comparison between the measured water content and the applicable purity specification or the water moisture limit in the process.

    Read more

    • The laboratory reports water content as a percent mass, calculated from the coulometric titration data.
    • The technician checks the type of organic liquid tested and determines that the liquid is within the allowed water content range of 0% to 2.0% mass.
    • The analyst documents the charge measured during electrochemical iodine generation and associates it with the sample batch being tested.
    • When interpreting the trace water contaminant results, the engineer takes into account the freshness of the reagents, exposure to moisture in the atmosphere, and injection technique.
    • Engineers use the measured water content values to support chemical purity verification and moisture-sensitive process monitoring decisions. 

    Link to ASTM E1064 

    FAQ

    Why does ASTM E1064 use coulometric titration instead of volumetric titration?
    The coulometric technique is suitable for water levels in organic liquids that are lower than those that could produce a small, imprecise volumetric Karl Fischer titrant volume, and it provides a direct measurement of the electrical charge without requiring the measurement of a titrant volume.
    No, this test method does not apply to liquefied gas products like LPG, butane, propane, or LNG, and there is a different applicable test method to determine water content in these product types.
    This test method is applicable to water content in the range from 0% to 2.0% by mass of the liquid organic chemicals, and is especially useful for detecting trace water content in liquid organic chemicals, rather than higher water content.

    Updated on October 1, 2026

    Davis Scott
    About Author
    Davis Scott is an Electrical and Electronics Engineer specializing in multidisciplinary validation, quality assurance, and comprehensive electro-mechanical testing.
    Know More
    Testimonials
    Real results
    Engineers trust us with what matters most
    Why Choose Matestlab?
    Matestlab is a unified testing services partner based in the USA. Matestlab offers raw material and product testing services through accredited, expert laboratories primarily based in the USA. Matestlab supports physical, thermal, chemical, electrical, mechanical, environmental, and characterization testing needs through its global network of 2200+ accredited testing laboratories. Matestlab's in-house materials science experts curate and design comprehensive testing programs matched to the sample material and specific testing need, with regulatory requirements and confidentiality in mind.

    If you are looking for a managed and trusted testing partner to achieve your research goals, feel free to schedule a meeting with us or send us a request at connect@matestlabs.com to learn more about our services. For a quick quote, call us at (302)244-7710. We are happy to serve you.
    Start Your Testing
    Project Today
    Define your requirements and get access to
    specialized laboratories ready to deliver results
    Partners with us
    Clients
    Vendors
    Process for testing
    • STEP 01

      You share your testing requirements

    • STEP 02

      You share your sample(s)

    • STEP 03

      We deliver your test reports

    Get your testing done

    Let us know your testing requirements and we will be right back with a solution.

      Let us root for each other. Collaborate to grow, expand, and accelerate our businesses.

      Partner with us