ASTM D6843 is a standard test method for characterizing bis-(triethoxysilylpropyl) sulfane-type silanes using gas chromatography. The method focuses on volatile components present in the silane rather than providing a complete chemical characterization of every component.
The scope includes:
- Material Type: Bis-(triethoxysilylpropyl)sulfanes used in rubber formulations.
- Analytical technique: Gas chromatography.
- Detector: Flame ionization detector (FID).
- Measured property: Volatile components and volatile impurities.
- Internal standard: Undecane is used as the internal standard.
- Result: Volatile impurities are reported in weight percent.
- Test limitations: The procedure measures volatile components and may not capture non-volatile or high-molecular-weight species.
- Applications: Evaluation of silane quality for rubber formulations.
- Units: SI units are the standard units of the method.
What are Bis-(Triethoxysilylpropyl)sulfanes?
Bis-(triethoxysilylpropyl)sulfanes are organosilane compounds used as coupling agents in rubber formulations. They help improve interactions between silica fillers and rubber polymers. These silanes can contain different sulfur-chain structures and may also contain volatile reaction products or impurities.
What Does ASTM D6843 Measure?
ASTM D6843 measures the amount of volatile components present in the silane sample. The chromatogram contains peaks corresponding to individual volatile components, and their peak areas are used to calculate the weight percentage of each component.
Components with retention times at or below the specified reference compound are treated as volatile impurities or volatile by-products. When the identity of an unknown volatile component is not known, the method applies the response factor of chloropropyltriethoxysilane (Cl-PTES) for calculation.
What are the Uses and Benefits of ASTM D6843 Testing?
ASTM D6843 provides useful information about the purity and consistency of silanes used in rubber formulations. Manufacturers can use the results to monitor raw materials and investigate changes in silane quality.
Key uses and benefits include:
- Quality control: Monitors the compositional consistency of silane products.
- Material evaluation: Compares the sulfane distribution of different silane batches or suppliers.
- Product development: Supports development and optimization of silane coupling agents for specific rubber formulations.
- Process evaluation: Helps identify changes in silane composition associated with manufacturing processes.
- Research: Provides controlled compositional data for silane chemistry studies.
- Regulatory compliance: Supports documentation of silane composition for safety and environmental reporting.
What Gas Chromatography Technique is Used in ASTM D6843?
The method uses gas chromatography with a flame ionization detector (FID). The silane sample is introduced into the GC, where its volatile components separate according to their interaction with the column and their volatility.
The FID detects the separated organic compounds and produces peaks in the chromatogram. The area of each peak provides the analytical signal used for calculating the concentration of the corresponding component.
A capillary column is typically used because it provides the separation required to distinguish volatile components in the silane mixture.
How is the Chromatographic Data Recorded?
The system can record detector response and retention times manually, through a chart or data logger, or with computerized equipment. For recorded data, the system must provide adequate resolution relative to the expected output signal. Computerized calculations are permitted when the system has been appropriately verified.
What Calibration is Required for ASTM D6843?
Calibration is important because the GC measures very small compositional differences. The detector should be calibrated using known concentrations of reference standards. Retention times should also be verified periodically.
The complete GC system should also be calibrated using at least one reference material with known composition. The calibration cycle should closely match the analytical conditions used during testing.
How is Data Collected in ASTM D6843?
The GC generates a chromatogram containing peaks for the separated components. The analyst records the retention time and peak area of each relevant compound.
The analysis is normally complete when the chromatographic baseline returns after the broad peak associated with higher sulfur-chain species. The method indicates that a typical run takes less than 30 min.